Arthur Holly Compton was awarded the Nobel Prize in 1927 for his work, published in 1923, of careful spectroscopic measurements of x-rays scattered at various angles by light elements. He found that x-rays scattered at larger angles had systematically larger wavelengths. He discovered that the observations were accounted for by considering the scattering as a collision between a single photon and a single electron in which energy and momentum are conserved. This effect now bears his name. The Compton effect demonstrates the essential duality of waves and particles in an especially clear way: Modeled as a particle (localized, having energy and momentum) one can apply conservation of energy and momentum to predict the relation between scattered X-ray energy and scattering angle. On the other hand, modeled as a wave, one can understand X-ray interference and diffraction phenomena.
* 1References | ||
* 21 Theory | ||
* 32 Experimental technique | ||
* 3.12.1 Overview | ||
* 3.22.2 PMT+NaI scintillator detector | * 3.3[2.3 Radioactive source ](#CopyofComptonScattering090519-2.3Radioactivesource) | * 3.42.4 Calibration sources |
* 3.52.5 Lab notebook considerations | ||
* 43 Experimental procedure | ||
* 4.13.1 Turn on and set the PMT high voltage | ||
* 4.23.2 View pulses on the scope | ||
* 4.33.3 Adjust the high voltage and gain | ||
* 4.43.4 Energy calibration | ||
* 4.4.13.4.1 In-software calibration | ||
* 4.4.23.4.2 Post-experiment calibration | ||
* 4.53.5 Collecting scattering data | ||
* 4.5.13.5.1 Considerations for data taking | ||
* 4.63.6 Preliminary analysis ("Day 2 Question") | ||
* 54 Analysis | ||
* 5.14.1 Energy calibration of pulse hight spectrum | ||
* 5.24.2 Scattering relationship | ||
* 5.34.3 Angular acceptance | ||
* 6Rubric | ||
* 6.1Autumn quarter | ||
* 6.2Winter quarter | ||
* 6.3Analysis Rubric | ||
* 6.3.1Prelab question | ||
* 6.3.2Lab notebook item 1 | ||
* 6.3.3Lab notebook item 2 | ||
* 6.3.4Day two question |
* A. A. Bartlett, //Am J. Phys//. **32**, 120 (1964). This paper is a historical review of the experiments that were later explained by Compton's discovery of the Compton effect. |
* A. H. Compton, Am. J. Phys. **29**, 817 (1961). Compton reviews the experimental evidence and the theoretical considerations that led to the discovery and interpretation of x-rays acting as particles. |
The more advanced theory dealing with the cross-section and absolute intensity of Compton scattering will not be addressed here. For more information, see the Klein-Nishina formula.
Consider the scattering of a gamma (photon) from a free electron as shown in Fig. 1.
{ ${/download/attachments/222397151/Scattering.png?version=1&modificationDate=1567719712000&api=v2}$ Figure 1: An incident gamma of energy E “collides” with an electron and scatters with energy E' at angle θ relative to the initial trajectory.
The energy of a gamma scattered by a free electron, {
$E'$ _ , _depends on the scattering angle, [Math Processing Error]θ {
$\theta$ , and the energy of the incident gamma, [Math Processing Error]E {
$E$ . It can easily be derived from the conservation of energy and momentum as
{![]() | (1) |
where { $mc^2 = 511 \mathrm{keV}$ is the rest energy of the electron. This is the relationship which you will test in the lab.
{ ${/download/attachments/222397151/IMG_20160825_160014869.jpg?version=1&modificationDate=1567719712000&api=v2}${
${/download/attachments/222397151/Compton_apparatus.png?version=1&modificationDate=1567719712000&api=v2}$
Figure 2: The Compton scattering apparatus.
The experimental apparatus is shown schematically in Fig. 2. A collimated beam of 662 keV {![]() |
By positioning the PMT+NaI detector at various angles about the scatterer, the energy of the Compton-scattered gammas can be determined from the resulting spectrum at each angle. To do this you will need to calibrate the pulse height axis of the spectrum using gammas of known energies.
A PMT coupled to a NaI crystal is a common detector used to measure the energy of gammas. A monoenergetic beam of gammas incident on a PMT+NaI detector will produce a characteristic spectrum of pulse heights. Understanding the details of this pulse height spectrum is necessary for analyzing your data. A detailed description of the pulse height spectrum produced by a PMT+NaI detector is given on the NaI Detector Physics and Pulse Height Spectra page. Make sure that you understand this description before the end of the first day of the experiment. |
A pair of 137Cs sources produce 662 keV gammas. These sources sit at the center of a lead pig to shield you from the radiation. The radiation emerges from the pig in a collimated beam aimed at the scatterer in the middle of the table. |
CAUTION: Do not place any part of your body in front of the open port of this source for an extended time. This source is on the order of 1000 times stronger than the plastic button sources used in other labs. (The activity is of the order of _milli_cuires rather than _micro_curies).
The “source” is actually two sources having strengths as follows:
These activities are nominal values only, as the activity will decay with time. (Cesium-137 has a half-life of 30.17 years.) When not in use, the pig is “closed” by a tungsten rod inserted into the exit aperture of the pig. A locking brass door holds the plug in place.
To calibrate the pulse height axis of the PHA, a set of small radioactive sources is provided. Sources include 241Am, 133Ba, 57Co, 137Cs, and 22Na, and should yield discernible gamma peaks with energies between 59.5 keV and 661.6 keV.
You need not consider energies above 662 keV when doing your calibration.
Energies and relative intensities of the calibration sources are available from the nuclear decay schemes. Note that these sources all have low activity so as to not overwhelm the detector with counts and cause charge pileup (also known as voltage sag.) |
<blockquote> NOTEBOOK: While the pig is closed you should sketch the layout of the experimental apparatus in your lab notebook, and important dimensions should be recorded. Of particular interest are the following: </HTML>
</blockquote></HTML>
Do not attempt to remove the NaI+PMT detector! Additional information about the apparatus geometry – including dimensions of several hard-to-measure quantities – is given in a separate PDF. You may consult this document, but be sure to verify measurements (and determine uncertainties) yourself, recording values in your notebook. | 3 Experimental procedure3.1 Turn on and set the PMT high voltageTo minimize electronic drifts, the power supply and pulse height analyzer should be turned on early and left on for the duration of the experiment. The pulse height-to-energy calibration can change significantly during the first half hour after the high voltage (HV) is first turned on.
The size of the pulses from the PMT scale with the magnitude of the high voltage applied to the tube. Later, we will adjust this voltage (and the amplifier gain) in order to take full advantage of the dynamic range of the PHA, but for now we can do some preliminary tests. 3.2 View pulses on the scopeThis is a good time to take a look at the PMT pulses on a scope. Before we send the signal through the PHA (which can be a bit of a black-box), it's good to understand the characteristics of the raw signal and check the behavior of the apparatus.
* Move the detector to zero degrees – directly in line with the source – and remove the aluminum rod scatterer.
The collimated beam from the source is now directly incident on the detector and we should have a very strong signal. Adjust the scope until you can observe the pulses.
If you are using an analog scope (the ones with the glowing green screen), then the screen will display not only the most recent trigger event, but many of the previous events which have not yet faded away. Since we can see many pulses at once, we can actually learn a bit about probabilities by comparing the intensity of different pulses.
3.3 Adjust the high voltage and gain
Now we wish to adjust the HV so that the full energy peak of the highest energy {
<blockquote> NOTEBOOK: Once you have finished setting the high voltage, save the resulting spectrum to the hard drive. It is recommended that you save all the spectra you collect in this experiment in both *.spu and *.tsv formats. It is good practice to record the filenames you use in the appropriate part of your lab notebook, along with enough information to know exactly what the spectra in the files represent. </HTML> NOTEBOOK:
</blockquote></HTML> *The purpose of the above exercise is to confirm that you have correctly identified the full energy peak associated with the 662 keV gammas. Mistakes are possible due to the presence of multiple gammas with overlapping features or because a high voltage or gain set was set too high or too low. In experimental work it is important to understand your data well enough to verify that things make sense. Otherwise you can waste time on something which is not what you are trying to study. If your calculated and measured values for the location of the Compton edge are consistent with one another then you can proceed with the experiment with increased confidence that you are on the correct path. If however these values are inconsistent, you should identify the problem before proceeding. This exercise is intended to be a 'quick & dirty' check; we are not looking for precise agreement. WARNING: Once you have set the high voltage and gain, DO NOT adjust either for the remainder of the experiment. 3.4 Energy calibration
3.4.1 In-software calibrationThe PHA software has a two- or three-point calibration feature which allows the x-axis of the display to change from channel number to energy. To calibrate in this way, one needs to know the true energy and the corresponding channel location for two (or three) gamma spectrum features.
3.4.2 Post-experiment calibrationThe in-software calibration method described above has several limitations.
Therefore, it is preferable to collect data in raw channel number and then, at home, fit the channel versus energy data to a function to find the conversion formula. NOTE: It does not hurt to do the two-point calibration described above even when you plan to do a better calibration later. When exporting the data in the *.tsv format, both the channel number and calibrated energy values are saved. In this way, you may use the rough calibration values from the software as a guide while in lab, but do the proper calibration when producing final plots and analysis.
It is suggested that you do a quick plot of the {
NOTE: The energy calibration is not necessarily constant over time. In fact, it is likely to shift a small amount, even under ideal circumstances. It is good practice to establish the stability of the calibration by repeating some measurements multiple times over the course of performing the experiment; in this way, you can quantify any changes over time. At a minimum, you should take calibration data at the beginning of your data collection and at the end on each day. By looking at how the calibration changes over the course of the experiment, you gain insight into an important source of systematic uncertainty in the data. Some factors which can cause the calibration of the detector to change over time include the following:
3.5 Collecting scattering data
NOTE: Choosing the centroid and uncertainty by moving your cursor around may arbitrary way to assign an uncertainty to a measured quantity, but uncertainties really are nothing more than the experimenter's judgement of how well they have measured a quantity; this is the simplest method. Another legitimate way to estimate the uncertainty in this measurement would be to estimate the location of the full energy peak with the cursor multiple times (say 10 or 20), average those values and look at the standard deviation of the average. One could also fit the full energy peak to a function like a Gaussian sitting on a background and obtain the value of the centroid and its uncertainty more quantitatively. For this experiment, fitting the peaks may mean a lot of extra work, but with only a small gain in the accuracy of the measurement. Using the cursor to estimate by eye gives an answer that is “good enough” because it turns out that other factors out of your control may dominate over the uncertainty in the peak locations. Use your judgement and justify your choices in the report. 3.5.1 Considerations for data takingIn this experiment you need to decide how many and at what angles to take data, as well as how best to determine the location of the center of the full energy peaks from the spectrum. When it comes to deciding on a data collecting strategy, you are faced with balancing trade-offs associated with the fact that you have a finite amount of time to do the experiment. How many angles should you make measurements at? Which angles should you use? How long do you spend collecting data at each angle? Learning how to reason your way to an appropriate answer to these questions is a big part of this course. Here are some things to consider as you plan your data collecting strategy.
CAUTION: Do NOT take data over night. Doing so is unnecessary in terms of getting an adequate number of counts, and the change in the calibration can be significant. 3.6 Preliminary analysis ("Day 2 Question")In completing the following task between Day 1 and Day 2, you may completely ignore uncertainties. Using your Day 1 calibration data, obtain a preliminary calibration of the detector channel-energy relationship. Then, select one of your Compton scattering spectra taken at any one fixed angle on Day 1, extract the channel location of the photopeak in the spectrum – you may do this approximately; we just want a rough estimate – and convert it to an energy for the scattered gamma using your preliminary calibration. Assuming that we know the incident gamma ray energy for the Cs-137 source, use the energy of the scattered gamma that you just found and the known scattering angle for selected spectrum to extract a rough value for the electron mass. Is it close to the literature value? Again, you do not need a statistically rigorous comparison here; a percentage difference is fine. A TA will check with you to discuss this calculation at the start of Day 2. 4 Analysis4.1 Energy calibration of pulse hight spectrumDetermine the energy calibration to be used for your analysis of the scattering data. You should have taken multiple sets of calibration data. Examine how stable the calibration was over the course of your experiment. Depending on how many calibration data sets you have, and how much the calibration changed over time, you have a couple of options.
However you decide to handle the calibration, you should justify your method. A table, plot and fit of the pulse height to energy calibration values you extract from the different calibration runs should be included. Include a discussion of the uncertainties in the calibration. If uncertainties had to be propagated through calculations, the formula for the propagation should be included. 4.2 Scattering relationship
Use your calibration(s) to determine the scattered {
After displaying the fit, discuss how well your data agree with the Compton scattering formula and evaluate – in depth – the results. 4.3 Angular acceptanceUsing the sketch and measurements of the geometry of your apparatus collected in Sec. 2.5, compute the _angular _acceptance of the detector (i.e. the full opening angle over which gammas scattered from the rod could enter the detector). For the report, discuss how this extended angular acceptance compares to your uncertainty in the angle position of the detector and discuss what affect a non-zero angular acceptance has on the data you collect. Note that this angular acceptance is not necessarily an explicit contribution to the uncertainty, but you should discuss its impact on how you collect data and what you measure. RubricWhen writing your report, consult the rubric and notes below for the appropriate quarter. Autumn quarter
Winter quarter
Analysis RubricIn writing your report it should be assumed that the reader is fully familiar with the contents of the wiki. Note that complete omission of a section will result in 0 points.
|